For Hydrocarbon-Free Air
Compact, regenerative hydrocarbon adsorbers capable of delivering up to 7.5 lpm of air with a hydrocarbon content of less than 1 ppm. Designed for GC-FID and emissions monitoring applications. Accessories include an inlet particulate filter, inlet coalescing filter, outlet filter-regulator assembly, moisture indicator, flow meter, surge tank and sound suppression kit. Standard Features:
Optional Features:
|
HYDROCARBON ADSORBERS - DESCRIPTION OF OPERATION
Ordinary compressed inlet air is passed through a charcoal bed where the hydrocarbons are adsorbed. The purified air then continues to the outlet side of the manifold and on to the application. A portion of the purified air is expanded through an orifice and directed back through an off-line charcoal bed, purging the accumulated hydrocarbons. Every 30 seconds, a solid state timer and two solenoid valves reverse the process. The result is a continuous supply of hydrocarbon-free air. |